HP Data Explorer 4 Series Manuale Utente Pagina 256

  • Scaricare
  • Aggiungi ai miei manuali
  • Stampa
Vedere la pagina 255
Chapter 5 Examining Spectrum Data
5-52 Applied Biosystems
5
Flexibility With the Peak Width parameter, determines the number
of points used to estimate the baseline amplitude at
regularly spaced points in the spectrum, but is not directly
proportional to the number of points used. Valid entries
are 0 to 1.0.
The default value of 0.5 works best for most applications.
A value closer to 0 reduces flexibility and provides a
smoother more generalized baseline correction (see
below).
A value closer to 1 provides more localized baseline
correction by using a larger number of points to estimate
the baseline (see below).
If the baseline rises adjacent to peaks after
baseline-correcting, decrease the Flexibility value and
baseline-correct again.
(continued)
Parameter Description/Specifies
Flexibility=1Flexibility=0
Vedere la pagina 255
1 2 ... 251 252 253 254 255 256 257 258 259 260 261 ... 446 447

Commenti su questo manuale

Nessun commento