HP E1312A and HP E1412A Multimeter Specifications 171Appendix A
Appendix A
HP E1312A and HP E1412A Multimeter Specifications
DC Characteristics
Accuracy Specifications
±
(% of reading + % of range) [1]
Function Range [3]
Test Current
or Burden
Voltage
24 Hour [2]
23°C ±1°C
90 Day
23°C ±5°C
1 Year
23°C ±5°C
Temperature
Coefficient
0°C - 18°C
28°C - 55°C
DC Voltage 100.0000mV
1.000000V
10.00000V
100.0000V
300.0000V
0.0030 + 0.0030
0.0020 + 0.0006
0.0015 + 0.0004
0.0020 + 0.0006
0.0020 + 0.0018
0.0040 + 0.0035
0.0030 + 0.0007
0.0020 + 0.0005
0.0035 + 0.0006
0.0035 + 0.0030
0.0050 + 0.0035
0.0040 + 0.0007
0.0035 + 0.0005
0.0045 + 0.0006
0.0045 + 0.0030
0.0005 + 0.0005
0.0005 + 0.0001
0.0005 + 0.0001
0.0005 + 0.0001
0.0005 + 0.0003
Resistance [4] 100.0000
1.000000k
10.00000k
100.0000k
1.000000M
10.00000M
100.0000M
1mA
1mA
100µA
10
µA
5
µA
500nA
500nA || 10M
0.0030 + 0.0030
0.0020 + 0.0005
0.0020 + 0.0005
0.0020 + 0.0005
0.0020 + 0.0010
0.0150 + 0.0010
0.3000 + 0.0100
0.0080 + 0.0040
0.0080 + 0.0010
0.0080 + 0.0010
0.0080 + 0.0010
0.0080 + 0.0010
0.0350 + 0.0010
0.8000 + 0.0100
0.0100 + 0.0040
0.0100 + 0.0010
0.0100 + 0.0010
0.0100 + 0.0010
0.0100 + 0.0010
0.0540 + 0.0010
0.8000 + 0.0100
0.0006 + 0.0005
0.0006 + 0.0001
0.0006 + 0.0001
0.0006 + 0.0001
0.0010 + 0.0002
0.0030 + 0.0004
0.1500 + 0.0002
DC Current 10.00000mA
100.0000mA
1.000000A
3.000000A
<0.1V
<0.7V
<1V
<2V
0.0050 + 0.0100
0.0100 + 0.0040
0.1000 + 0.0060
0.7000 + 0.0200
0.0500 + 0.0200
0.0500 + 0.0050
0.1300 + 0.0100
0.7200 + 0.0200
0.0700 + 0.0200
0.0700 + 0.0050
0.1500 + 0.0100
0.7200 + 0.0200
0.0050 + 0.0020
0.0060 + 0.0005
0.0060 + 0.0010
0.0060 + 0.0020
DC:DC Ratio 100mV to 300V (Input Accuracy) + (Reference Accuracy)
Input Accuracy = accuracy specification for the HI-LO input signal.
Reference Accuracy = accuracy specification for HI-LO reference
input signal (Sense HI-LO input terminals).
NOTE:
Autorange is used for the reference signal regardless of the range set
for the HI-LO input signal. The 10V range is the highest range available
for the reference signal and the highest range the multimeter will
autorange to for measuring the reference signal.