Chapter 4, Troubleshooting
Questionable Measurement Results
Sensitivity tests: random data not selected for signal source
For receiver sensitivity measurements, check to see that the CDMA Generator is
set to send random data. In the CDMA Generator Menu Selection screen, the
Generator Mode should be set to Data and the Data Source should be set to
Random.
Performance in additive noise: signal source not set for Eb/No
For this test, noise is added to the channel using the AWGN generator. Verify that
the Generator Mode is set to Eb/No and that the Eb/No Level is set to an
appropriate level.