HP 132A Manuale Utente Pagina 92

  • Scaricare
  • Aggiungi ai miei manuali
  • Stampa
  • Pagina
    / 330
  • Indice
  • SEGNALIBRI
  • Valutato. / 5. Basato su recensioni clienti
Vedere la pagina 91
Chapter 3 Programming Your Universal Counter for Remote Operation
Status Reporting
3-24 Programming Guide
Standard Event Status Register Group
Figure 3-8. Standard Event Status Reporting
Standard Event Status Register
The Standard Event Status Register contains bits that monitor specific IEEE
488.2-defined events as shown in Figure 3-8.
Use *ESR? to read this register.
Use *ESR? or *CLS to clear this register.
Table 3-5 lists the Standard Event Status Register bits and briefly describes each
bit.
Table 3-5. Standard Event Status Register
BIT WEIGH
T
SYMBOL DESCRIPTION
0 1 OPC Operation Complete
1 (RQC) Not used because this instrument cannot
request permission to become active
IEEE 488.1 controller-in-charge.
2 4 QYE Query Error
3 8 DDE Device-Specific Error
4 16 EXE Execution Error
5 32 CME Command Error
6 (URQ) Not used because this instrument does not
define any local controls as User Request
controls.
7 128 PON Power On
Vedere la pagina 91
1 2 ... 87 88 89 90 91 92 93 94 95 96 97 ... 329 330

Commenti su questo manuale

Nessun commento